Vulnerability Identified in Philips IntelliBridge EC40/80 Hubs

By | November 15, 2019

A vulnerability has been identified in the Philips IntelliBridge EC40/80 hub which could allow an attacker to gain access to the hub and execute software, modify files, change the system configuration, and gain access to identifiable patient information.

Philips IntelliBridge EC40/80 hubs are used to transfer medical device data from one format to another, based on set specifications. The hub does not alter the settings or parameters of any of the medical devices to which it connects.

The vulnerability could be exploited by an attacker to capture and replay a session and gain access to the hub. The flaw is due to the SSH server running on the affected products being configured to allow weak ciphers.

The vulnerability would only require a low level of skill to exploit, but in order to exploit the flaw an attacker would need to have network access. The flaw – CVE-2019-18241 – has a CVSS v3 base score of 6.3 out of 10 – Medium severity.

The flaw was reported to Philips by New York-Presbyterian Hospital’s Medical Technology Solutions team, and under its responsible vulnerability disclosure policy, Philips reported the vulnerability to the DHS Cybersecurity Infrastructure Security Agency.

The vulnerability is present in all versions of the EC40 and EC80 hubs and will be addressed in a new release, which will not be available until the end of Q3, 2020.

Until Philips issues the new release, users of the affected hubs have been advised to implement the following mitigation measures to reduce the potential for exploitation.

  1. Only operate the hub within Philips authorized specifications, using Philips approved software, configurations, system services, and security configurations
  2. There is no clinical requirement for these devices to communicate outside the Philips clinical network. The devices should be logically or physically separated from the hospital network.
  3. Users should block access to the SSH port. SSH is not meant to be used for clinical purposes, only for product support.
  4. Use a long and complex SSH password and make sure password distribution is controlled to ensure SSH is used via physical access only.